EQMS-P-8100
Uniphos Online COD Analyzer
(Potentiometric Method)
- Municipal Water
- Industrial Process Fluids
- Wastewater and Effluent Water
- Ground and Surface water
- COD analysis by direct method, which involves oxidation of sample by potassium dichromate and Measurement by potentiometric titration.
- Ten times reduction in chemicals and cost per analysis as compared to traditional method
- A fully automated system with a industry accredited Program Logic controller (PLC) and Human Machine Interface (HMI)
- Built in diagnostic features for reagent level alarm and the component malfunctions
- Data communication compatible with required SPCB/CPCB norms
- Measurement Method: Potentiometric Measurement
- Measuring Range: 0-500 mg/l / 50-2000 mg/l
- Resolution: 1 mg/l / 15 mg/l
- Accuracy: > 100 mg/l, ±10% / >250 mg/l, ±10%
- < 100 mg/l, ±20 mg/l / < 250 mg/l, ±40 mg/l
- Digestion Temperature: 180o C for 20 minutes
- Analysis Time: 60 minutes
- Power Requirement: 230V AC/ 60 Hz
- Data Communication: RS-485 & 4-20 mA
- Dimensions: H=1200mm, L=400mm, B=600mm.
- Weight: 70 kg Approx.